NASA will be performing network maintenance on Wednesday, November 20, 2024, between 11:00 AM EST and 1:00 PM EST (16:00 to 18:00 UTC). There will be two periods of about five minutes of disruption. During these peroids, the ILRS website will be unavailable during the outages. If you experience problems outside this window, contact support-cddis@nasa.gov.

close window
sitelog navigation structure General Site Log Meteorological Data LAGEOS Performance Satellite Data Info

Mt. Stromlo: Site Log

Instructions and Update Site Log

You can use the hyperlinks to quickly get to the section of interest or you can scroll down to read them all.

Section: 0. Form
Section: 1. Identification of the Ranging System Reference Point (SRP)
Section: 2. Site Location Information
Section: 3. General System Information
Section: 4. Telescope Information
Section: 5. Laser System Information
Section: 6. Receiver System
Section: 7. Tracking Capabilities
Section: 8. Calibration
Section: 9. Time and Frequency Standards
Section: 10. Preprocessing Information
Section: 11. Aircraft Detection
Section: 12. Meteorological Instrumentation
Section: 13. Local Ties, Eccentricities, and Collocation Information
Section: 14. Local Events Possibly Affecting Computed Position
Section: 15. On-Site, Point of Contact Agency Information
Section: 16. Responsible Agency (if different from 15.)
Section: 17. More Information

 
       ILRS Site and System Information Form
       International Laser Ranging Service
 
  0.   Form
 
       Prepared by (Full Name)    : Jim Steed and Chris Moore
       Preparer E-mail            : jim.steed@ga.gov.au,
                                  : stromlo_slr@eos-aus.com
       Date Prepared              : 2003-03-10
       Report Type                : UPDATE
       Site Log Format Version    : 2.0
       Site Log Revision          : 
 
 
  Back to top
 
 
  1.   Identification of the Ranging System Reference Point (SRP)
 
       Site Name                  : Mount Stromlo
       IERS DOMES Number          : 50119S001
       CDP Pad ID                 : 7849
       Subnetwork                 : WPLTN
       Description                : AZ EL INTERSECT
       Monument Description       : N.A.
       Monument Inscription       : N.A.
       Mark Description           : N.A.
       Date Installed             : 1998-06-19
       Date Removed               : 
       Geologic Characteristic    : BEDROCK
       Additional Information     : Bedrock is FRESH IGNEOUS.
                                  : Installation date is of first system 
                                  : acceptance. Initial date for system is 
                                  : 1997-11-03.
 
 
  Back to top
 
 
  2.   Site Location Information
 
       City or Town               : Canberra
       State or Province          : Australian capital territory
       Country                    : Australia
       Tectonic Plate             : Australian-Indian
       Local Time (UTC +/- hours) : 
       Approximate Position
         X coordinate          [m]: -4467063.9
         Y coordinate          [m]: 2683034.5
         Z coordinate          [m]: -3667007.2
         Latitude            [deg]: 35.3161 S
         Longitude           [deg]: 149.0099 E
         Elevation             [m]: 804.1
       Survey Method              : 
       Date Measured              : 
       Additional Information     : 
 
 
  Back to top
 
 
  3.   General System Information
 
  3.01 System Name                : Stromlo
       4-Character Code           : STRL
       CDP System Number          : 80
       CDP Occupation Number      : 01
       Eccentricity to SRP (if Not Identical With SRP)
         North                 [m]: 
         East                  [m]: 
         Up                    [m]: 
         Date Measured            : N.A.
       Date Installed             : 1997-11-03
       Date Removed               : 2003-01-18
       Additional Information     : The Stromlo SLR station was destroyed by fire
                                  : along with many buildings of the nearby 
                                  : Mt Stromlo Atronomical Observatory during
                                  : the 18th January 2003 bush fires that also 
                                  : destroyed over 500 houses in the suburbs of
                                  : Canberra.
 
 
  Back to top
 
 
  4.   Telescope Information
 
  4.01 Receiving Telescope Type   : Confocal paraboloid
         Aperture              [m]: 0.75
         Mount                    : AZ-EL
       Date Installed             : 
       Date Removed               : 
       Xmitting Telescope Type    : Confocal paraboloid
         Aperture              [m]: 0.75
       Date Installed             : 
       Date Removed               : 
       Tracking Camera Type       : ICCD
         Model                    : MX12P/8X43
         Manufacturer             : ADIMEC
         Field of View       [deg]: 0.0167
         Minimum Magnitude   [mag]: 7
       Date Installed             : 
       Date Removed               : 
       Transmit/Receive Path      : COMMON COUDE
       Transmit/Receive Switch    : POLARIZER
       Max Slew Rate Az    [deg/s]: 10
       Max Slew Rate El    [deg/s]: 10
       Max Used Tracking Rate Az  : 4
       Max Used Tracking Rate El  : 3
       Telescope Shelter          : SEALED DOME
       Daylight Filter Type       : Andover 11622 Fabry-Perot
       Dayl. Filt. Bandwidth  [nm]: 0.15
       Adjustable Attenuation     : YES
       Adjustable Field of View   : 
       Transmit Efficiency        : 0.41
       Receive Efficiency         : 0.41
       Date Installed             : 1998-06-19
       Date Removed               : 2003-01-18
       Additional Information     : Transmit/Receive Switch: 
                                  :     Dichroic Mirror plus SHG;
                                  :     Ssubstitued "Polarizer";
                                  :     Is that correct;
                                  :     rlr. 07/12/2019
 
 
  Back to top
 
 
  5.   Laser System Information
 
  5.01 Laser Type                 : ND:YAG
       Model                      : 
       Manufacturer               : 
       Number of Amplifiers       : 2
       Primary Wavelength     [nm]: 1064
       Primary Maximum Energy [mJ]: N.A.
       Secondary Wavelength   [nm]: 532.1
       Secondary Max. Energy  [mJ]: 50
       Xmit Energy Adjustable     : NO
         Additional Information   : 
       Pulse Width (FWHM)     [ps]: 30
       Max. Repetition Rate   [Hz]: 100
       Fullw. Beam Divergence  ["]: 7 - 60
         Theoretical Beam Div     : 
         Measured Beam Div        : 
           Method Used            : 
           Additional Information : 
       Final Beam Diameter     [m]: 0.75
       Eyesafe                    : NO
       Eyesafe Standard           : ANSI Z136.1 - 1993
       Polarization on Exit       : 
       Laser Fire Control         : 
       Laser Energy Monitor       : 
       Start Diode Type           : 
         Manufactorer             : 
         Model                    : 
         Quantum Efficiency    [%]: 
         Rise/Fall Time       [ps]: 
         Date Installed           : 
         Date Removed             : 
         Signal Processing        : 
           Manufacturer           : 
           Model                  : 
           Date Installed         : 
           Date Removed           : 
           Additional Information : 
       Date Installed             : 1998-06-19
       Date Removed               : 2003-01-18
       Additional Information     : Microlaser oscillator at 100 MHz, microjoule
                                  : Region. Pulses selected and amplified in a
                                  : Regenerative amplifier, then fed to a double
                                  : Pass power amplifier
                                  : Primary Maximum Energy: 
                                  :     not used for laser ranging
                                  : 
 
 
  Back to top
 
 
  6.   Receiver System
 
  6.01.01 Primary Chain
          Wavelength          [nm]: 532
          Detector Type           : CSPAD
            Manufacturer          : PESO Consulting
            Model                 : ASY01643
            Quantum Efficiency [%]: 20
            Nominal Gain          : 
            Rise Time         [ps]: 100
            Jitter (Single PE)[ps]: 22
            Field of View Diam ["]: 8 - 32
            Date Installed        : 1999-09-01
            Date Removed          : 
          Amplifier Type          : 
            Manufacturer          : 
            Model                 : 
            Gain                  : 
            Bandwidth             : 
            Date Installed        : 
            Date Removed          : 
            Additional Information: 
          Signal Processing       : Time Walk Compensated
            Manufacturer          : Graz
            Model                 : 
            Date Installed        : 1999-09-01
            Date Removed          : 
            Amplitude Measurement : YES
            Return-Rate Controlled: NO
            Mode of Operation     : Single to Multiple Photons
            Additional Information: 
          Time of Flight Observ.  : EVENT
            Manufacturer          : EOS
            Model                 : MRCS V.4
            Resolution        [ps]: 0.7
            Precision         [ps]: 10
            Date Installed        : 1998-06-19
            Date Removed          : 2003-01-18
          Additional Information  : 80 MHz reference oscillator derived from 
                                  : OCXO. Mode of operation depends on amplitude 
                                  : filtering application which depends on 
                                  : received return rate (i.e. post processing
                                  : configurable)
                                  : Rise Time: <100
 
  6.02.01 Secondary Chain
          Wavelength          [nm]: 
          Detector Type           : 
            Manufacturer          : 
            Model                 : 
            Quantum Efficiency [%]: 
            Nominal Gain          : 
            Rise Time         [ps]: 
            Jitter (Single PE)[ps]: 
            Field of View Diam ["]: 
            Date Installed        : 
            Date Removed          : 
          Amplifier Type          : 
            Manufacturer          : 
            Model                 : 
            Gain                  : 
            Bandwidth             : 
            Date Installed        : 
            Date Removed          : 
            Additional Information: 
          Signal Processing       : 
            Manufacturer          : 
            Model                 : 
            Date Installed        : 
            Date Removed          : 
            Amplitude Measurement : 
            Return-Rate Controlled: 
            Mode of Operation     : 
            Additional Information: 
          Time of Flight Observ.  : 
            Manufacturer          : 
            Model                 : 
            Resolution        [ps]: 
            Precision         [ps]: 
            Date Installed        : 
            Date Removed          : 
          Additional Information  : 
 
  6.03.01 Tertiary Chain
          Wavelength          [nm]: 
          Detector Type           : 
            Manufacturer          : 
            Model                 : 
            Quantum Efficiency [%]: 
            Nominal Gain          : 
            Rise Time         [ps]: 
            Jitter (Single PE)[ps]: 
            Field of View Diam ["]: 
            Date Installed        : 
            Date Removed          : 
          Amplifier Type          : 
            Manufacturer          : 
            Model                 : 
            Gain                  : 
            Bandwidth             : 
            Date Installed        : 
            Date Removed          : 
            Additional Information: 
          Signal Processing       : 
            Manufacturer          : 
            Model                 : 
            Date Installed        : 
            Date Removed          : 
            Amplitude Measurement : 
            Return-Rate Controlled: 
            Mode of Operation     : 
            Additional Information: 
          Time of Flight Observ.  : 
            Manufacturer          : 
            Model                 : 
            Resolution        [ps]: 
            Precision         [ps]: 
            Date Installed        : 
            Date Removed          : 
          Additional Information  : 
 
 
  Back to top
 
 
  7.   Tracking Capabilities
 
  7.01 Satellites
         Very Low Alt (<400 km)   : YES
         Low Altitude (400-2000)  : YES
         LAGEOS/LARES             : YES
         GLONASS                  : YES
         Etalon                   : YES
         GNSS                     : YES
         Geosynchronous           : 
         Moon                     : NO
       Ave. Pass Switch Time   [s]: 60
       Average values for LAGEOS
         Single Shot RMS      [mm]: 10
         # of Obs per NP          : 100
         Use of Semi-trains       : NO
         # of Semi-train Tracks   : N.A.
       Range Gate Width       [ns]: 200 - 5200
       Beam Pointing Accuracy  ["]: 3.5
       Angle Encoder Resolution["]: 0.3
       Min. Tracking Elev.   [deg]: 20
       Laser Polarization @emitter: 
       Operation
         Months per Year          : 12
         Days per Week            : 7
         Hours per Day            : 24
         Staff per Shift          : 1
         System Shared With       : R&D/SPACE DEBRIS
         Time Allocated to SLR [%]: 67 - 100
         Remotely Controllable    : YES
       Tracking Restrictions
         Go/No-go flag            : 
           Satellite Verification : 
           Date Verified          : 
         Maximum elevation        : 
           Satellite Verification : 
           Date Verified          : 
         Pass segments            : 
           Satellite Verification : 
           Date Verified          : 
         Schedule segments        : 
           Satellite Verification : 
           Date Verified          : 
         Maximum laser power limit: 
           Satellite Verification : 
         Date Verified            : 
       Date First Applicable      : 1998-10-29 HH:MM UT
       Date Last Applicable       : 2003-01-18 HH:MM UT
       Additional Information     : Staff per Shift: 
                                  :     1 half the time, 
                                  :     0 in autonomous operation
 
 
  Back to top
 
 
  8.   Calibration
 
  8.01 Calibration Type           : PRE+POST
       Target Location            : EXTERNAL
       Target Type                : CORNER CUBE open-face
       Target Structure           : METAL POLE in protective metal tube
       Number of Targets          : 
       Target Distance(s)      [m]: 69.5346
         Date Measured            : 1999-06-01
         Accuracy             [mm]: 1
         Verification             : Weekly ranging to 4 ground targets
                                  : plus real time internal target. 
       Return-Rate Controlled     : 
       Eye-Safe Method            : 
         Additional Information   : 
       Mode of Operation          : SINGLE/MULTI PHOTON
       Calibration performed      : 180
       Single Shot RMS        [mm]: 4.3
       Edit Criterion 1st Chain   : ITERATIVE 2.5 SIGMA
       Edit Criterion 2nd Chain   : N.A.
       Application of Cal Data    : LINEAR FIT
       Date Installed             : 1999-01-15
       Date Removed               : 2003-01-18
       Additional Information     : Occasionally use just PRE calibration with 
                                  : application of AVERAGE
 
  8.02 Calibration Type           : REALTIME
       Target Location            : INTERNAL
       Target Type                : CORNER CUBE open-face
       Target Structure           : N.A.
       Number of Targets          : 
       Target Distance(s)      [m]: 1.29
         Date Measured            : 1998-06-30
         Accuracy             [mm]: 2
         Verification             : Weekly ranging to 4 ground targets plus
                                  : real Time internal target.
       Return-Rate Controlled     : 
       Eye-Safe Method            : 
         Additional Information   : 
       Mode of Operation          : SINGLE/MULTI PHOTON
       Calibration performed      : 
       Single Shot RMS        [mm]: 
       Edit Criterion 1st Chain   : ITERATIVE 2.5 SIGMA
       Edit Criterion 2nd Chain   : N.A.
       Application of Cal Data    : AVERAGE
       Date Installed             : 1998-06-19
       Date Removed               : 1999-01-15
       Additional Information     : Facility is quasi-available and will be
                                  : restored as principlal calibration method
                                  : as soon as possible.
 
 
  Back to top
 
 
  9.   Time and Frequency Standards
 
  9.01.01 Frequency Standard Type : Oven Controlled Crystal Oscillator
          Model                   : XL-DC 151-602-254
          Manufacturer            : TrueTime
          Short Term Stab.  [e-12]: 100
          Long Term Stab.   [e-12]: 1
          Time Reference          : GPS
          Synchronization         : GPS
          Epoch Accuracy      [ns]: 150
          Date Installed          : 1998-06-19
          Date Removed            : 2003-01-18
          Additional Information  : TrueTime or HP depending upon which
                                  : GPS Timing Receiver is used. OCXO GPS
                                  : disciplined with low phase noise output
                                  : options.
 
  9.01.02 Frequency Standard Type : Cesium
          Model                   : 5071A
          Manufacturer            : HP
          Short Term Stab.  [e-12]: 5
          Long Term Stab.   [e-12]: 0.01
          Time Reference          : GPS
          Synchronization         : GPS
          Epoch Accuracy      [ns]: 150
          Date Installed          : 1998-10-29
          Date Removed            : 2003-01-18
          Additional Information  : Backup and monitoring standard
 
  9.02.01 GNSS Timing Rcvr Model  : XL-DC 151-602-254
          Manufacturer            : TrueTime
          Date Installed          : 1998-06-19
          Date Removed            : 2003-01-18
          Additional Information  : 
 
  9.02.02 GNSS Timing Rcvr Model  : 58503
          Manufacturer            : HP
          Date Installed          : 1998-10-29
          Date Removed            : 2003-01-18
          Additional Information  : Backup receiver
 
  9.03.01 One-way System Calibration
          Date performed          : 
          Direction Calibrated    : 
          Additional Information  : 
 
 
  Back to top
 
 
 10.   Preprocessing Information
 
 10.01 On-site NP Generation      : YES
       Data Screening             : IRV+POLYNOMIAL
       Return-Rate Rejection      : 
       Edit Criterion 1st Chain   : 2.5 SIGMA
       Edit Criterion 2nd Chain   : N.A.
       Upload interval            : DAILY
       Date First Applicable      : 1999-07-01 HH:MM UT
       Date Last Applicable       : 
       Additional Information     : Automatic process ("AUTOPROC")
 
 10.02 On-site NP Generation      : YES
       Data Screening             : IRV+POLYNOMIAL
       Return-Rate Rejection      : 
       Edit Criterion 1st Chain   : MANUAL + 2.5 SIGMA
       Edit Criterion 2nd Chain   : N.A.
       Upload interval            : DAILY
       Date First Applicable      : 1998-06-19 HH:MM UT
       Date Last Applicable       : 1999-06-30 HH:MM UT
       Additional Information     : Manual process ("PROFITS")
 
 
  Back to top
 
 
 11.   Aircraft Detection
 
 11.01 Detection Type             : EYESAFE LASER
       Manufacturer/Provider      : 
       Automatic Laser Shutoff    : 
       Alarms Recorded            : 
       Hazard Distance        [km]: 
       Date Installed             : 1998-03-01
       Date Removed               : 2003-01-18
       Additional Information     : 1570 nm OPO conversion from Infinity Nd:YAG
                                  : Full-rate synchronous, fired 290 us before
                                  : Primary emission. Energy <30mJ, width 3.2ns
 
 
  Back to top
 
 
 12.   Meteorological Instrumentation
 
 12.01.01 Pressure Sensor Model   : PTB200A
          Manufacturer            : Vaisala
          Recording Interval      : EVERY 5 SECONDS
          Accuracy          [mbar]: 0.05
          Height Diff to SRP   [m]: +0.21 
          Distance from SRP    [m]: 
          Date Installed          : 1998-06-19 HH:MM UT
          Calibration Interval    : DAILY/YEARLY
          Date Removed            : 2003-01-18 HH:MM UT
          Additional Information  : Calibrated daily against local reference,
                                  : and every 2 years against Meteorological
                                  : Bureau standards.
 
 12.01.02 Pressure Sensor Model   : Mercury
          Manufacturer            : FORTIN
          Recording Interval      : DAILY
          Accuracy          [mbar]: 0.5
          Height Diff to SRP   [m]: -4.43
          Distance from SRP    [m]: 
          Date Installed          : 1999-08-03 HH:MM UT
          Calibration Interval    : 
          Date Removed            : 2003-01-18 HH:MM UT
          Additional Information  : Used as back-up barometer.
                                  : Readings are manual.
 
 12.02.01 Temp Sensor Model       : HMP350
          Manufacturer            : Vaisala
          Recording Interval      : EVERY 5 SECONDS
          Accuracy         [deg C]: 0.1
          Date Installed          : 1998-06-19 HH:MM UT
          Calibration Interval    : see additional comments
          Date Removed            : 2003-01-18 HH:MM UT
          Additional Information  : Factory calibration.
 
 12.02.02 Temp Sensor Model       : 9858253 Mercury
          Manufacturer            : Dobbie
          Recording Interval      : Daily
          Accuracy         [deg C]: 0.1
          Date Installed          : 1998-06-19 HH:MM UT
          Calibration Interval    : 
          Date Removed            : 2003-01-18 HH:MM UT
          Additional Information  : Back-up thermometer. Readings
                                  : are manual.
 
 12.03.01 Humidity Sensor Model   : HMP350
          Manufacturer            : Vaisala
          Recording Interval      : EVERY 5 SECONDS
          Accuracy       [% rel h]: 0.5
          Date Installed          : 1998-06-19 HH:MM UT
          Calibration Interval    : 
          Date Removed            : 2003-01-18 HH:MM UT
          Additional Information  : Integral with primary temperature sensor.
 
 
  Back to top
 
 
 13.   Local Ties, Eccentricities, and Collocation Information
 
 13.01 Collocated Permanent Geodetic Systems
 
 13.01.01.01 GNSS                 : IGS
             Date Installed       : 1996-06-08
             Date Removed         : 2003-01-18
             Additional Info      : 
 13.01.01.02 GNSS                 : IGLOS
             Date Installed       : 1998-09-17
             Date Removed         : 2003-01-18
             Additional Info      : 
 13.01.02.01 DORIS                : IDS
             Date Installed       : 1998-10-22
             Date Removed         : 2003-01-18
             Additional Info      : 
 13.01.03.01 PRARE                : NO
             Date Installed       : 
             Date Removed         : 
             Additional Info      : 
 13.01.04.01 VLBI                 : NO
             Date Installed       : 
             Date Removed         : 
             Additional Info      : There is a VLBI system at Tidbinbilla 
                                  : 9.6 km away
 13.01.05.01 Gravimeter           : SUPERCONDUCTING
             Date Installed       : 1997-01-01
             Date Removed         : 
             Additional Info      : This instrument belongs to Japanese 
                                  : National Astronomical Observatory.
                                  : It is in a basement of the Astronomical
                                  : Observatory, about 600 meters from the 
                                  : SLR system. 
 13.01.06.01 Other SLR            : 
             Date Installed       : 
             Date Removed         : 
             Additional Info      : 
 
 13.02.01.01 Local Ties from the SRP to Other Monuments or Systems on Site
 
          Monument Name           : STR1
          Instrumentation Type    : GPS
          Instrumentation Status  : PERMANENT
          DOMES Number            : 50119M002
          CDP Number              : 
          Differential Components (ITRS)
            dx                 [m]: -38.7148 +- 0.0006
            dy                 [m]: 4.9877 +- 0.0008
            dz                 [m]: 57.5162 +- 0.0005
            Date Measured         : 1999-06-01
            Determined by         : AUSLIG
          Date Installed          : 1998-06-15
          Date Removed            : 
          Additional Information  : Intersection of vertical axis of 5/8 inch
                                  : Whitworth threaded steel stub with horizontal
                                  : plane coinciding with top of mounting plate
                                  : to which steel stub is attached. This
                                  : mounting is on top of the North calibration
                                  : pillar, 183 mm above a similar mounting used
                                  : for the standard SLR calibration target.
                                  : Monument inscription - AU052
 
 13.02.02.01 Local Ties from the SRP to Other Monuments or Systems on Site
 
          Monument Name           : STR0
          Instrumentation Type    : FIDUCIAL MONUMENT
          Instrumentation Status  : PERMANENT
          DOMES Number            : 50119M001
          CDP Number              : 
          Differential Components (ITRS)
            dx                 [m]: -10.7687 +- 0.0005
            dy                 [m]: -22.6353 +- 0.0008
            dz                 [m]: -0.7403 +- 0.0005
            Date Measured         : 1999-06-01
            Determined by         : AUSLIG
          Date Installed          : 1998-01-01
          Date Removed            : 
          Additional Information  : Intersection of vertical axis of 5/8 inch
                                  : Whitworth threaded stainless steel spigot
                                  : with top of stainless steel pillar plate set
                                  : Into top of the fundamental survey pillar at
                                  : Stromlo SLR station.
                                  : GLONASS/GPS antenna mounted on spigot.
                                  : Monument Inscription - AU045
                                  : Fundamental Pillar
 
 13.02.03.01 Local Ties from the SRP to Other Monuments or Systems on Site
 
          Monument Name           : DORIS
          Instrumentation Type    : DORIS
          Instrumentation Status  : PERMANENT
          DOMES Number            : 50119S002
          CDP Number              : 
          Differential Components (ITRS)
            dx                 [m]: -4.4647 +- 0.0007
            dy                 [m]: -4.3871 +- 0.0012
            dz                 [m]: 3.3729 +- 0.0006
            Date Measured         : 1999-06-01
            Determined by         : AUSLIG
          Date Installed          : 1998-10-22
          Date Removed            : 2003-01-18
          Additional Information  : Intersection of the vertical axis of Starec
                                  : type antenna with plane coinciding with the
                                  : painted reference line midway up the antenna.
 
 13.03.01.01 Eccentricities Between Other Monuments on Site
 
          From: Monument Name     : 
                DOMES Number      : 
                CDP Number        : 
          To:   Monument Name     : 
                DOMES Number      : 
                CDP Number        : 
          Differential Components (ITRS)
            dx                 [m]: 
            dy                 [m]: 
            dz                 [m]: 
            Date Measured         : 
            Determined by         : 
          Additional Information  : 
 
 
  Back to top
 
 
 14.   Local Events Possibly Affecting Computed Position
 
 14.01 Date                       : 2003-01-18 HH:MM UT
       Event                      : Fire storm
       Additional Information     : SLR station destroyed along with GPS, Doris 
                                  : and Glonass monitoring equipment. Monuments 
                                  : STR0 and STR1 not detroyed but will require 
                                  : re-furbishment and re-survey once the site 
                                  : is re-established. 
 
 14.02 Date                       : 2003-07-04 HH:MM UT
       Event                      : Facility replacement begins
       Additional Information     : The Australian Government, through Geoscience
                                  : Australia (formerly AUSLIG) signed a 
                                  : contract with EOS Space Systems Pty Limited 
                                  : (EOS) to rebuild the Stromlo SLR facility.
 
 14.03 Date                       : 2003-09-09 HH:MM UT
       Event                      : Operation of facility
       Additional Information     : Another contract signed with EOS Space
                                  : Systems Pty Limited (EOS) for the operation 
                                  : and maintenance of the Geoscience Australia
                                  : facilty. The building is already completed
                                  : awaiting the installation of the dome, 
                                  : telescope and laser. Engineering data is 
                                  : expected to be available in December 2003, 
                                  : with full operational status in early 2004.
 
 
  Back to top
 
 
 15.   On-Site, Point of Contact Agency Information
 
       Agency                     : Electro Optic Systems Pty. Ltd.
       Mailing Address            : Locked Bag 2, PO
                                  : Queanbeyan  NSW  2620
                                  : Australia
       Website                    : 
       Primary Contact
         Contact Name             : Dr Christopher Moore
         Telephone (primary)      : +61 2 6287-2953
         Telephone (secondary)    : +61 2 6298-8000
         Fax                      : +61 2 6287-2951
         E-mail                   : stromlo_slr@eos-aus.com
       Secondary Contact
         Contact Name             : Mr Ron Thompson
         Telephone (primary)      : +61 2 6298-8018
         Telephone (secondary)    : 
         Fax                      : +61 2 6299-7687
         E-mail                   : ronthompson@eos-aus.com
       Tertiary Contact
         Contact Name             : 
         Telephone (primary)      : 
         Telephone (secondary)    : 
         Fax                      : 
         E-mail                   : 
       Additional Information     : 
 
 
  Back to top
 
 
 16.   Responsible Agency (if different from 15.)
 
       Agency                     : National Mapping Division, 
                                  : Geoscience Australia
       Mailing Address            : PO Box 2
                                  : Belconnen  ACT  2616
                                  : Australia
       Primary Contact
         Contact Name             : Jim Steed
         Telephone (primary)      : +61 2 6201-4347
         Telephone (secondary)    : +61 2 6201-4201
         Fax                      : +61 2 6201-4366
         E-mail                   : jim.steed@ga.gov.au
       Secondary Contact
         Contact Name             : Mr John Manning
         Telephone (primary)      : +61 2 6201-4352
         Telephone (secondary)    : +61 2 6201-4201
         Fax                      : +61 2 6201-4366
         E-mail                   : john.manning@ga.gov.au
       Additional Information     : 
 
 
  Back to top
 
 
 17.   More Information
 
       URL for More Information   : www.eos-aus.com
       Hardcopy on File
         Site Map                 : YES
         Site Diagram             : YES
         Horizon Mask             : NO
         Monument Description     : YES
         Site Pictures            : YES
       Additional Information     : 
 
 
  Back to top